Giacomo Bertocchi > Radiant heaters GB9 series
русengita

Radiant heaters GB9 series

The radiant warmers series GB9R treat the hypothermia of newborns, warming them by invisible infrared light, controllable and safe, the baby remains completely accessible for vision and manipulations.

  • The Infant Warmers GB 9R can be coupled to any cot and working plan
  • The irradiated energy can be controlled or manually, or in automatic mode, supporting the preset skin temperature of a baby
  • Basement with pirouetting castors.
  • Height adjustment of the radiant system
  • The heating system is constituted by irradiating elements in refractory material with invisible light (in order not to modify the chromatism of the new-born).
  • Large area of uniform heating energy distribution
  • The irradiation is invisible and doesn’t distort the colour of newborn’s skin.

THE MAIN FEATURES OF THE GB9R models:

  • The heating system is constituted by irradiating elements in refractory ceramic with invisible light
  • Visualisation of the energy delivered from the warmer through 10 LEDs
  • Manual or automatic (depends of the model) electronic regulation of the radiant thermal energy
  • Audible signal to be silenced every 15 minutes, when the delivered power exceeds the 20% of the total power
    Possibility to set up the new-born’s skin temperature from 36 to 38 C°
  • Audible and visual alarms for:
    - Power failure
    - High and low temperatures: +1C° or -1C° compared to set up value (only for ISC models)
    - power failure
    - Acoustic signal every 15 minutes, if operated in manual mode, to call the operator to check up the baby.

The GB9R panel can have only manual control of heating energy, or both manual and automatic control systems ( the models with automatic control are indicated with ISC prefix). These models are equipped with automatic thermocontroller and baby’s skin probe.

The real and preset skin temperatures are displayed.
The temperature can be preset in the range of 36 - 38°C.
The mode of operation can be selected by the operator.

The ISC models have the following addition alarms:

  • High temperature – +1°C as compared to preset value;
  • Low temperature - -1°C as compared to preset value.

 

The GB9R panels are available in two versions: :

Model GB9RN –
With mobile stand with 4 castors, 2 of them – with brakes, one – antistatic.
This panel can be combined with any cot, swaddling table, working plane.

Model GB9RM –
With 2 holders to be attached at the wall over the working plane (for example, table for washing and dressing the newborns).

TECHNICAL DATA

Power requirements : 220 V 50 Hz
Max input power : 450 VA
Weight : 26 Kg
GB9RN 26 kg
GB9RM 19 kg
Castors diameter
mod. GB9RM 60 mm

BERTOCCHI S.r.l. ELETTROMEDICALI
Via De' Berenzani 8 a/b 26100 Cremona Italia
Tel +39 0372 471000
Fax +39 0372 471533
e-mail: info@bertocchielettromedicali.com   |   www.bertocchielettromedicali.com